Surgical Asepsis and Assisting with Surgical Procedures


Surgical Asepsis and Assisting with Surgical Procedures


Learning Objectives



1. Define, spell, and pronounce the terms listed in the vocabulary.


2. Apply critical thinking skills in performing the patient assessment and patient care.


3. Define the concepts of aseptic technique.


4. Explain the differences among sanitization, disinfection, and sterilization.


5. Summarize tips for improving autoclave techniques.


6. Demonstrate how to wrap instrument packs for autoclave sterilization.


7. Explain the types and uses of sterilization indicators.


8. Summarize the correct methods of loading, operating, and unloading an autoclave.


9. Demonstrate how to operate an autoclave.


10. Summarize common minor surgical procedures.


11. Detail the medical assistant’s role in minor office surgery.


12. Perform a skin prep for surgery.


13. Perform a surgical hand scrub.


14. Outline the rules for setting up and maintaining a sterile field.


15. Open a sterile pack to create a sterile field.


16. Transfer sterile instruments and pour solutions into a sterile field.


17. Put on sterile gloves without contaminating them.


18. Demonstrate how to assist with a minor surgical procedure and suturing.


19. Summarize postoperative instructions and care of wounds.


20. Demonstrate how to remove sutures and the technique for removing surgical staples.


21. Explain the process of wound healing.


22. Properly apply dressings and bandages to surgical sites.


23. Conduct patient education in aseptic technique and surgical procedures.


24. Discuss the legal and ethical concerns regarding surgical asepsis and infection control.


Vocabulary


cicatrix Early scar tissue that appears pale, contracted, and firm.


dehiscence The separation of wound edges or rupture of a wound closure.


infection Invasion of body tissues by microorganisms, which then proliferate and damage tissues.


sterilization Complete destruction of all forms of microbial life.


Scenario


Melissa Gelbart, CMA (AAMA), works for a dermatologist, Dr. Susan Armstrong, who frequently performs minor surgical procedures in the office. Melissa was hired to work as an administrative medical assistant at the front desk, but one of the clinical medical assistants has unexpectedly quit, and the office manager has offered Melissa her position. Melissa is excited about this opportunity, but she also is concerned about her skill level in sterile procedures. At least she is familiar with a number of the patients, most of the staff, and the types of outpatient surgeries performed in the facility. Surgical asepsis and assisting with surgery were her favorite topics when she was in medical assisting school. However, before she can assist with surgeries, Melissa must demonstrate her ability to set up a sterile field without contaminating the site. She also must show that she can apply a sterile dressing properly and change sterile bandages.


While studying this chapter, think about the following questions:



• What are the crucial steps Melissa must follow to set up and maintain a sterile field?


• How does an autoclave work and what are the important rules to remember when preparing surgical trays for the autoclave and correctly operating the machine?


• How will Melissa know whether surgical trays processed in the autoclave are actually sterile?


• What techniques must Melissa follow to prepare for and assist with a surgical procedure?


• What are common surgical procedures performed in an ambulatory care facility?


• What is the medical assistant’s role in preparing the patient, equipment, and room for a surgical procedure?


• Why is it important that Melissa understand and be prepared to answer patients’ questions about the process of wound healing?


• What bandaging techniques should Melissa be prepared to perform?


Asepsis is the condition of being free of infection or infectious material. Medical asepsis is the destruction of organisms after they leave the body. The principles of medical asepsis are implemented to prevent reinfection of a patient and cross-infection of another patient or ourselves. To prevent cross-contamination, potential microorganisms and pathogens must be isolated by following standard blood and body fluid precautions and by disinfecting or sterilizing objects as soon as possible after they become contaminated. As discussed in Chapter 27, medical asepsis is the process of either reducing the number of pathogens or destroying them; this creates an environment that is clean but not sterile (free of microorganisms).


Surgical asepsis is the complete destruction of organisms on instruments or equipment that will enter the patient’s body. This technique is mandatory for any procedure that invades the body’s skin or tissues, such as surgery. Everything that comes in contact with the patient must be sterile, including surgical gowns, drapes, and instruments, in addition to the gloved hands of the surgeon and surgical assistants. Anytime the skin or a mucous membrane is punctured or pierced, as in venipunctures or injections, aseptic techniques must be practiced. Urinary catheterizations, biopsies, and dressing changes on open wounds are performed using sterile technique.


A medical assistant must develop an inner sense of sterile procedures. It is important that these techniques be performed on such a routine basis that they become an unbreakable habit. Conscientious attention must be given to sterilizing all items at all times. Frequent checking and rechecking of procedures helps ensure that they are effective and are used without any “breaks” in technique. Single-use, disposable items offer the best method of infection control, and they are being used more frequently in medical offices. However, when disposable equipment is used, the assistant must know the specific disposal guidelines for contaminated instruments and supplies.


Sterilization


Before an instrument or piece of equipment can be used in a surgical procedure, it first must be sanitized, then disinfected, and finally sterilized to remove all forms of microorganisms. Sanitization and disinfection were described in Chapter 27. It is essential that you understand these two concepts, so review them if necessary before learning sterilization methods.


Instruments and other items used in office surgery, examination, or treatment must be carefully cleaned before proceeding with the steps of disinfection or sterilization. Sanitization is the cleansing process that reduces the number of microorganisms to a safe level as dictated by public health guidelines. This cleansing process removes debris such as blood and other body fluids from instruments or equipment. Blood and debris must be removed so that later disinfection with chemicals or sterilization with steam, heat, or gases can penetrate to all the instrument’s surfaces (see Procedure 27-5). The procedure should be completed immediately after the instruments are used. If this is not possible, rinse the used items under cold water immediately after the surgical procedure and place them in a low-sudsing, rust-inhibiting, enzyme-containing, detergent solution. Never allow blood or other substances that can coagulate to dry on an instrument.


The medical assistant should always wear gloves while performing sanitization (thick utility gloves if the instruments have sharp or pointed edges) to prevent possible personal contamination with potentially infectious body fluids that may be present on the articles being cleaned. When you are ready to sanitize instruments, drain off the soak solution and rinse each instrument in cold, running water. Separate the sharp instruments from the others, because metal instruments may damage the cutting edges, and sharp instruments may damage the other instruments or injure you. Clean all sharp instruments at one time, when you can concentrate on preventing injury to yourself. Open all hinges and scrub serrations and ratchets with a small scrub brush or toothbrush. Rinse the instruments in hot water, then check them carefully for proper working order before they are disinfected or sterilized. The items should be hand dried with a towel to prevent spotting.


Disinfection is the process of killing pathogenic organisms or of rendering them inactive. However, it is not always effective against spores, the tubercle bacilli, and certain viruses. Disinfectant chemicals may kill microbes within a short time but are usually very hard on instruments. Some chemicals, such as Cidex, are effective enough to kill all organisms, but the usual immersion time for these sterilants is 10 hours or longer. Many types of disinfecting agents are available and have varying degrees of effectiveness. It is important to follow the manufacturer’s guidelines on the proper use of each product and also to understand the product’s advantages and disadvantages and the possible sources of error.


To ensure proper sterilization for surgical aseptic procedures, an area (usually a utility room) should be set aside in each office for just this purpose. The area should be divided into two sections, one dirty and one clean. The dirty section is used for receiving contaminated instruments and other materials at the conclusion of surgical procedures. This area should have a sink, receiving basins, proper cleaning agents, brushes, utility gloves, autoclave wrapping paper or cloth, autoclave envelopes and tape, sterilizer indicators, and disposable gloves. Designated biohazardous waste containers are needed for gloves worn when handling contaminated items. Personal protective equipment (PPE) for autoclave procedures includes:



The clean section of the utility room should be reserved for receiving the sterile items after they have been removed from the sterilizer. Clear, clean plastic bags in which to store sterile packs may be kept in the clean area. Both areas should be spotlessly clean and well organized. Sterilization can be achieved by moist heat in an autoclave, by gas, or with chemicals. Most medical offices use the autoclave method. A written sterilization procedure should be in place for each workplace.



57-1


Critical Thinking Application


The office manager told Melissa she needs to review the office policy and procedures manual on surgical supplies and sterilization methods. Why is this important before Melissa starts performing sterilization procedures? What information in this manual would be most important to Melissa as she starts this new position? Why?


Autoclave


Steam under pressure in the autoclave (Figure 57-1) is the best method of sterilization, because it kills all pathogens and spores. Pressurized steam is fast, convenient, and dependable. The pressure allows for heat higher than the boiling point, and when combined with moisture, these two factors create a very effective mechanism for killing all microorganisms. When steam is admitted into the autoclave chamber, it simultaneously heats and wets the object, coagulating the proteins present in all living organisms. When the cycle is complete and the chamber has cooled, the steam condenses and explodes the cells of microorganisms, thus destroying them. To be effective, the steam moisture must come in contact with all surfaces being sterilized. Steam under pressure is capable of much faster penetration of fabrics and textiles than dry heat, but its use has definite limitations if the proper techniques are not followed.



The recommended temperature for sterilization in an autoclave is 121° to 123° C (250° to 255° F). Unwrapped items should be sterilized for 20 minutes, small wrapped items for 30 minutes (Table 57-1), and large or tightly wrapped items for 40 minutes. Processing time starts after the autoclave reaches normal operating conditions of 121° C (250° F) and 15 pounds per square inch (psi) pressure.



TABLE 57-1


Sterilization Chart*




































































































ARTICLE METHOD TEMPERATURE (° F) TIME
Gauze, small, loosely packed Autoclave 250° 30 min
Gauze, large, loosely packed Autoclave 270° 30 min
Gauze, small, tightly packed Autoclave 250° 40 min
Gauze, large, tightly packed Autoclave 270° 40 min
Gauze, tightly packed Dry heat 320° 3 hr
Gauze, loosely packed Dry heat 320° 2 hr
Glass syringes in tubes Autoclave 250° 30 min
Glass syringes in muslin Dry heat 320° 1 hr
Instruments on tray, muslin under and over Dry heat 320° 1 hr
Instruments on tray, muslin under and over Autoclave 250° 15 min
Solutions in flasks with gauze plug Autoclave 250° 30 min
Glassware unwrapped Dry heat 320° 1 hr
Glassware wrapped Autoclave 250° 30 min
Petroleum jelly, 1-ounce jar Dry heat 340° 1 hr
Petroleum jelly, 2-ounce jar Dry heat 320° 2 hr
Petroleum gauze in instrument tray Dry heat 320° 150 min
Powder, 1-ounce jar Dry heat 320° 2 hr
Powder, small glove packs Autoclave 250° 15 min


Image


*Remember always to place a sterilization indicator in areas where there is doubt the steam will penetrate. Do not assess effectiveness by chamber pounds per square inch; a thermometer and sterilization indicator are the reliable methods of judging a killing temperature.


The three basic autoclave cycles are as follows:



• Gravity (“fast exhaust”) cycle: This cycle is used to sterilize stainless steel instruments, glassware, and so on. The autoclave fills with steam and is held at a set temperature for a set period. When the cycle is complete, a valve opens and the chamber rapidly returns to atmospheric pressure. Drying time may be added to the end of the cycle. This is the cycle most often used in the physician’s office setting.


• Liquid (“slow exhaust”) cycle: This cycle is used to prevent sterilized liquids from boiling. Steam is exhausted slowly at the end of the cycle, allowing the liquids to cool.


• Prevacuum cycle: This cycle is used for porous materials. The chamber is partially evacuated before the introduction of steam for greater steam penetration; this is not available on all machines.


Incorrect operation of an autoclave may result in superheated steam. If steam is brought to too high a temperature, it is literally dried out, and the advantage of a higher heat is diminished. Wet steam is another cause of incomplete sterilization. Wet steam results from failing to preheat the chamber, which causes excessive condensation in the interior of the chamber. Condensation is necessary, but too much prevents the sterilization process from being completed properly. It can be compared with taking a hot shower in a cold bathroom, which results in heavily steamed mirrors, walls, and towels. If packs become too saturated to dry during the drying cycle, the packs pick up and absorb bacteria from the air or any surface on which they are placed after removal from the autoclave. Placing cold instruments in a hot chamber also increases condensation. Other causes of wet steam include opening the door too wide at the end of the cycle or allowing a rush of cold air into the chamber. Overfilling the water reservoir may produce this same effect.


The main cause of incomplete sterilization in the autoclave is the presence of residual air. Without the complete elimination of air, an adequately high temperature cannot be reached. Air and steam do not mix. Because air is heavier than steam, it pools wherever possible. One tenth of 1% (0. 1%) residual air trapped around an instrument prevents complete sterilization. This is especially dangerous in older autoclaves that do not have a chamber thermometer separate from the pressure gauge. Adequate chamber pressure does not guarantee a proper chamber temperature. Table 57-2 provides tips for improving autoclave techniques.



TABLE 57-2


Tips for Improving Autoclave Techniques








































PROBLEM CAUSES CORRECTION
Damp linens Clogged chamber drain; goods removed from chamber too soon after cycle; improper loading Remove strainer; free openings of lint. Allow goods to remain in sterilizer an additional 15 min with door slightly open. Place packs on edge; arrange for least possible resistance to flow of steam and air.
Stained linens Dirty chamber Clean chamber with Calgonite solution; never use strong abrasives, such as steel wool; rinse thoroughly after cleaning.
Corroded instruments Poor cleaning; residual soil; exposure to hard chemicals (e.g., iodine, salt, and acids); inferior instruments Improve cleaning; do not allow soil to dry on instruments; sanitize first. Do not expose instruments to these chemicals; if exposure occurs, rinse immediately. Use only top-quality instruments.
Spotted or stained instruments Mineral deposits on instruments; residual detergents from cleaning; mineral deposits from tap water Wash with soft soap and detergent with good wetting properties. Rinse instruments thoroughly with distilled water.
Instruments with soft hinges or joints Corrosion or soil in joint; instrument parts out of alignment Clean with warm, weak acid solutions (10% nitric acid solution); rinse thoroughly. Have instrument realigned by qualified instrument repair professional.
Ebullition, or caps that blow off solutions Too rapid exhausting of chamber Use slow exhaust, cool liquids, or turn autoclave off and let cool on its own; that is, let the pressure drop at its own rate.
Steam leakage Worn gasket; door closes improperly Replace gasket; reopen door and shut carefully; have serviced if unable to close door properly.
Chamber door does not open Vacuum in chamber (check chamber pressure gauge) Turn on controls to starting steam pressure; wait until equalized, then vent and open door.

Wrapping Materials


Maintenance of sterility depends completely on the wrapper and method of wrapping (Procedure 57-1). The wrapping material must be permeable to steam but impervious to contaminants. Acceptable wrapping materials for autoclaving should be made of a substance that allows the steam to penetrate while preventing pathogens from entering during storage and handling. A wrapper should not be used if it is torn or has a hole in it. Clean muslin, disposable autoclave paper, and polypropylene bags are examples of autoclave instrument wraps (Figure 57-2).



Procedure 57-1


Prepare Items for Autoclaving


Wrap Instruments and Supplies for Sterilization in an Autoclave


GOAL: To place dry, checked, sanitized, and disinfected supplies and instruments inside appropriate wrapping materials for sterilization and storage without contamination.


EQUIPMENT and SUPPLIES



Procedural Steps



1. Sanitize your hands. Collect and assemble already sanitized and disinfected items to be wrapped. Gloves may be worn.


2. Place the wrapping material on a clean, flat surface.


3. Place the item (or items) diagonally at the approximate center of the wrapping material. Make sure the size of the square is large enough for the items (Figure 1).


image


PURPOSE: Each of the four corners must fold over and completely cover the items, with a few extra inches of overlap for folding.


4. With the squares that are cloth fabric, use two pieces if the cloth is single layered; follow the manufacturer’s recommendation when using commercial autoclave wrapping paper.
PURPOSE: To ensure sterility until the sterile item is needed for use.


5. Open any hinged instruments. If the instrument is sharp, its teeth or tip should be shielded with cotton or gauze.
PURPOSE: To prevent puncture of the package or injury to the operator.


6. If the package is to contain several items, place a commercial sterilization indicator inside the package at the approximate center.
PURPOSE: To ensure that the autoclave is reaching effective levels of heat and pressure.


7. Bring up the bottom corner of the wrap and fold back a portion of it.
PURPOSE: This folded-back flap is the only part of each wrapper corner that can be touched when a sterile package is opened (Figure 2).


image

8. Repeat the above step with each corner, making sure to turn back a portion each time (Figures 3 and 4).


image

image

9. Fold the last flap over (Figure 5).


image

10. Secure with autoclave tape (Figure 6).


image

11. Secure with autoclave tape and label the package with the date, including the year, contents, and your initials (Figure 7).


image


PURPOSE: So that staff members will know what is in the pack at a later date, whether the shelf life has expired (expiration date), and who performed the task. As a general rule, most office-autoclaved packs are considered sterile (usable) for up to 28 days.



Wrapping Instruments


The method used to wrap instruments for autoclave sterilization must allow the pack to be opened without becoming contaminated. The rules for protecting package contents include the following:



• Inspect muslin wrappers for holes before each use and discard if any holes are found.


• Wrap all hinged instruments in the open position to allow full steam penetration of the joint.


• Place a gauze sponge around the tips of sharp instruments to prevent them from piercing the wrapping material.


• If a number of instruments are to be placed on a stainless steel tray for wrapping, first place a double-folded towel on the tray, then position the instruments. This helps to protect them.


• Polypropylene is a plastic capable of withstanding autoclaving but is resistant to heat transfer. Therefore, materials in a polypropylene pan take longer to autoclave than the same materials in a stainless steel pan.


• When using sterilizing bags, insert the jaws of the instruments first to ensure that the grasping end of the instrument can be reached easily when the bag is opened.


• Indicate on the wrapper what is in the package or label it with a code. This code should correspond with a list of instruments that are stored with the pack after sterilization.


• Label each pack according to the instrument contents, sterilization date, and your initials. Use a permanent marker, never a ballpoint pen.


• Whether you are wrapping one item or many items together on a tray as a surgical pack, the procedure is the same; be sure the wrapper is large enough to cover the items to be sterilized.



57-2


Critical Thinking Application


Melissa is processing instruments and trays when she notices that one of her co-workers never inspects a muslin wrapper before wrapping a pack. What is the significance of Melissa’s observation? How should she handle this situation? Why?


Sterilization Indicators


Sterilization is achieved only when steam reaches the optimum temperature for a designated length of time and has penetrated to the center of the articles. Sterilization indicators must be used routinely to determine whether all microorganisms have been destroyed. The two basic types of sterilization indicators are chemical indicators (autoclave tape) and biologic indicators (bacterial spore strips).


Chemical Sterilization Indicators.

Autoclave tape, a commonly used sterilization indicator, contains a chemical dye that changes color when exposed to steam (Figure 57-3). The tape is not an absolute indication that the proper sterilization time, temperature, and steam have been maintained; it merely indicates that a high temperature was reached while the article was in the autoclave. The strip must completely change color (colors vary by manufacturer) or reveal the word “autoclaved” to ensure effective operation. The main function of autoclave tape, besides holding the wrapping material together or closing a sterilization bag, is to verify that the package has been autoclaved.



Biologic Sterilization Indicators.

The facility should have a policy for how frequently the autoclave is tested using biologic methods. One type, a spore strip indicator, contains a temperature-sensitive dye that changes color when the proper combination of steam, temperature, and time has been achieved. An indicator strip should be placed in the center of the largest pack that typically would be autoclaved in the facility to determine the accuracy of the autoclave and autoclave procedures. Test indicator kits are available that use ampules of Bacillus stearothermophilus, which is destroyed at 121° C (250° F). On completion of the cycle, the ampule is sent to the laboratory for analysis of any type of microbial growth, which would indicate that the autoclave is not sterilizing properly.


Quality Assurance Records for Office Sterilization


Every office should have specific protocols to follow for quality-assurance evaluations of the autoclave. This is done at specified intervals, depending on the volume and frequency of autoclave use. A log must be kept of the type of control test done, when it was performed, and the testing results. If the testing results indicate that sterilization was inadequate, a report must be made and filed. The report should identify the nature of the problem and how and when it was corrected. The report also should contain proof of correction by indicating the date and time of a first, subsequent, and successful sterilization run.


Loading the Autoclave


Prepare all packs and arrange the load in a way that allows maximum circulation of steam and heat (Procedure 57-2). Articles should be resting on their edges and should not be crowded. Placing the packs in stainless steel racks prevents packing of the autoclave too tightly. Jars, bottles, and trays must be wrapped and placed on their sides if they are to be used to store sterile items. Covers on jars and containers should be put to one side or left open to allow steam to penetrate. Extreme care must be taken not to contaminate jars when replacing their lids after autoclaving.



Procedure 57-2


Perform Sterilization Procedures


Operate the Autoclave


GOAL: To sterilize properly prepared supplies and instruments using the autoclave.


EQUIPMENT and SUPPLIES



Procedural Steps


NOTE: The specific instructions for operating an autoclave may vary based on the model number and manufacturer. Refer to the instructions that accompany the autoclave to be sure the appropriate steps are followed.



1. Check the water level in the reservoir and add distilled water as necessary.
PURPOSE: Too much or too little water may alter the effectiveness of the equipment. Tap water leaves lime deposits in the chamber.


2. Turn the control to “Fill” to allow water to flow into the chamber. The water flows until you turn the control to its next position. Do not let the water overflow.


3. Load the chamber with wrapped items, spacing them for maximum circulation and penetration.
PURPOSE: To ensure sterilization of all items.


4. Close and seal the door.
PURPOSE: The door must be closed, or the heated water in the chamber evaporates.


5. Turn the control setting to “On” or “Autoclave” to start the cycle.


6. Watch the gauges until the temperature gauge reaches at least 121° C (250° F) and the pressure gauge reaches 15 pounds (lb) of pressure.
PURPOSE: The proper temperature and pressure must be reached before sterilization can begin.


7. Set the timer for the desired time.


8. At the end of the timed cycle, turn the control setting to “Vent.”
PURPOSE: This releases the steam and pressure. The water at the bottom of the chamber drains back into the reservoir.


9. Wait for the pressure gauge to reach zero.


10. Standing behind the autoclave door, carefully open the chamber door ¼ inch.
PURPOSE: To allow steam to escape faster. Be careful to prevent accidental burns.


11. Leave the autoclave control at “Vent” to continue releasing heat.
PURPOSE: To dry the items faster.


12. Allow complete drying of all articles.


13. Using heat-resistant gloves, remove the items from the chamber and place the sterilized packages on dry, covered shelves or open the autoclave door and allow the items to cool completely before removal and storage.


14. Turn the control knob to “Off” and keep the door slightly ajar.
PURPOSE: To allow the inside of the autoclave to dry completely.


Instruments may be autoclaved unwrapped if they do not need to be sterile when used later. For example, although vaginal speculums do not need to be sterilized for use (the vagina is a body cavity that is naturally open to the external environment), they must be sanitized, disinfected, and sterilized to prevent cross-contamination among patients. They can be placed unwrapped on a perforated stainless steel tray in the autoclave and then stored in a clean area for future use.


Unloading Guidelines


When the autoclave’s sterilization cycle is complete, release the pressure according to the manufacturer’s guidelines. Once the pressure gauge reads “0,” stand back from the door and, with heat-resistant gloves, open the door approximately ¼ inch. Allow the load to dry for at least 15 minutes (this time varies according to the type of autoclave and the size of the load). Capillary attraction is the action that draws moisture through the surface of materials. Packs can act like a sponge, attracting outside moisture and microorganisms. Touching a wet pack allows microorganisms on your hands to penetrate the wrappings, making the contents of the pack nonsterile. Dry, wrapped packs may be removed with clean, dry hands, but it is safer to wear heat-resistant gloves to reduce the possibility of burns from the hot instruments inside the packs. If possible, allow all packs to cool in the autoclave with the door open. Place the packs on a dry, dust-free surface inside an enclosed cupboard or drawer for storage. Do not place the packs on cold surfaces, because hot packs may cause condensation, and moisture will contaminate the contents.


Guidelines for unloading an autoclave include the following:



Shelf Life of Sterilized Packs.

Each office has its own guidelines for the shelf life of sterile packs. Generally, muslin and autoclave paper packs are considered sterile for up to 28 days from the date of sterilization. Polypropylene autoclave bags are sterile for up to 6 months from the sterilization date. All sterile packs should be stored on dry, dust-free, covered shelves or in drawers. Fabric wrappers must be inspected for holes and laundered after each use. A damaged pack or a broken seal renders the package nonsterile; spills of any fluid onto a package also contaminate it. When a pack is no longer sterile for any reason, including the expiration date, the contents must be reprocessed as if the pack had been used for surgery. The contents must be sanitized, disinfected, wrapped, and sterilized as usual.



57-3


Critical Thinking Application


Melissa discovers a number of packages of paper-wrapped sterile instruments that have no dates on them. The indicator tape shows that they have been autoclaved. What should she do with these packs? Why?


Gas Sterilization


A variety of gas sterilizers are available. Each has its own very specific operating guidelines to ensure operator safety. Because of the long processing times, the very specific requirements for gas ventilation established by the Occupational Safety and Health Administration (OSHA), and the hazards of reproductive organ damage and cancer associated with gas sterilization, it is unrealistic to use gas sterilization in the physician’s office.


Chemical Sterilization


In the medical office, chemical sterilization is used for instruments that cannot be exposed to the high temperatures of steam sterilization. The sterilizing chemical solution must be mixed exactly according to the instructions on the bottle. The solution must be marked with the date of preparation and expiration. Materials to be sterilized must be submerged in this chemical bath with a closed lid for 8 hours or longer. Items are removed with sterile forceps and must be rinsed with sterile water to remove all traces of the chemical before the items are used on a patient. Removed items are then dried with a sterile towel. You must avoid skin contact with the sterilizing solution because it is very caustic.


Surgical Procedures


Common surgical procedures that are routinely performed in the primary care office include suturing, cyst removal, incision and drainage (I&D) of abscesses, and collection of biopsy specimens. The medical assistant should be proficient in explaining each of these procedures to the patient, preparing the patient and the room, assisting the physician with the surgery, and applying a sterile dressing and bandage after the procedure is finished.


Each surgical procedure requires appropriate skin preparation and draping with a fenestrated drape, also called an eye sheet. This is a surgical drape with an opening in the center. The size of the opening depends on the size of the surgical field. The opening is placed directly over the surgical site after the site has been suitably prepared (or “prepped,” as it is called in healthcare practice). A minor surgery tray is opened, and a sterile field is created on a Mayo instrument stand. Sutures, scalpel, and any other instruments needed are added to the field, according to the surgeon’s preference. Have a local anesthetic ready, also according to the physician’s preference.


After achieving suitable local anesthesia, the physician opens the skin with an incision. If a cyst is being removed, the physician dissects around it and usually tries to “deliver” it from the wound intact. If the procedure is an I&D, foul matter will start oozing from the wound immediately after the skin is incised. The wound is drained completely and flushed with copious amounts of sterile saline solution. A drain may be placed in the wound and left for several days. If the procedure is a biopsy, a small amount of tissue is removed and placed in a specimen container with preservative. The specimen container must be carefully labeled with the appropriate patient information, the date, and specifics about the specimen type and location. It then is sent to the laboratory, where it is examined microscopically for changes or abnormalities.


Electrosurgery


Electrosurgery is also known as electrocautery. An electrosurgical unit (ESU) uses high-frequency current to cut through tissue and coagulate blood vessels. A small probe with an electric current running through it is used to cauterize (burn or destroy) the tissue. When the electric current comes in contact with tissue and blood cells, they are vaporized, producing carbon and steam. This process seals blood vessels, minimizing cellular oozing and bleeding. Electrosurgery may be used to destroy granulations and small polyps.


Necessary components are the ESU’s power source, the grounding cable and pad, and the active electrode (a pencil-like instrument with a tip and cord). Tips are disposable and are used according to the type of procedure performed. The two most commonly used tips are the needle and flat designs.


Holding the pencil-like instrument, the surgeon touches the tissue with the tip and activates the electric current with a switch on the instrument or a foot pedal. The electric current is delivered to the tissues, and tissue is vaporized at the site of contact.



Laser Surgery


Laser is an acronym for light amplification by stimulated emission of radiation. Because a laser beam is so small and precise, it can be used to safely treat specific tissue with minimal damage to surrounding tissues and limited scar formation. Lasers were first used in medicine to treat diseases of the retina, and they now are used for many procedures, including excision of lesions, cauterization of blood vessels, removal of warts or moles, and cosmetic surgical procedures.


Several types of lasers are used, including the carbon dioxide, yttrium-aluminum-garnet (YAG), and pulsed dye lasers. Each laser has a specific use. The color of the laser light beam is directly related to the type of surgery performed.


A medical assistant must be specially trained to operate a laser before assisting with laser surgery. Laser equipment requires very careful handling, care, and maintenance. Laser light destroys tissue and can harm the patient, the physician, and you if handled improperly. The medical assistant should complete a full laser safety program before assisting in laser procedures. Once trained, the medical assistant’s role during laser surgery includes:



Microsurgery


Microsurgery involves the use of an operating microscope to perform delicate surgical procedures. One of its major uses is in ophthalmologic surgery. It also is used in otologic, rhinologic and sinus, laryngologic, neurosurgical, microvascular, gynecologic, and genitourinary procedures. A medical assistant must acquire a basic knowledge of the operation and care of a microscope before becoming qualified to assist in these types of procedures.


The basic components of an operating microscope are the light source, eyepieces (also called the oculars), lenses, and cord. Accessory pieces include assistant and observer lenses, cameras, video recorders, television monitors, and printers. These are all valuable for documentation and teaching purposes. Disposable sterile drapes and handle covers are used on the microscope during surgical procedures.


Surgical microscopes are expensive, delicate instruments that require extreme care in handling and cleaning. All lenses and cords should be carefully inspected before and after each use.


Endoscopic Procedures


An endoscope is a medical device consisting of a miniature camera mounted on a flexible tube with an optical system and a light source that is used to examine the area inside an organ or cavity. Many types of endoscopes are used, and they are named according to the organs or areas they are used to explore, such as the urinary bladder, bronchus, larynx, colon (Figure 57-4), stomach, uterus, abdomen, and various joints. Small instruments can be used to take samples of suspicious tissues through the endoscope.



Direct visualization with an endoscope is used for diagnostic purposes or to perform surgical procedures. Endoscopes may be rigid (e.g., laparoscope or hysteroscope), semirigid, or flexible (colonoscopes, bronchoscopes, gastroscopes). All are delicate and expensive and require extreme care in handling to protect them from damage.


Accessory equipment used with endoscopes includes fiberoptic light cables and light source; irrigators for solution instillation and suction; and a camera, monitor, printer, and video recorder. The fiberoptic light cable consists of hundreds of glass fibers. It is important to protect it from being bent, dropped, kinked, squashed, or smashed. The light source can become very hot and must be kept out of contact with the patient, the physician, the staff, and any flammable material, such as surgical draping. All equipment must be checked before and after use. Always follow the manufacturer’s recommendations for use, care, and maintenance of equipment.


Cryosurgery


Cryosurgery involves the use of a very-low-temperature probe to destroy tissue by freezing it on contact. The probe’s temperature usually is below −20° C (4° F). This cold temperature is achieved by circulating liquid nitrogen through the tip of the probe. A local anesthetic usually is administered before cryosurgery. Cryosurgery is used to treat cancers of the skin, prostate, liver, pancreas, and kidney. In many situations, cryosurgery is less invasive than traditional surgery and therefore generally has fewer associated complications. Cryosurgery often is performed in an office setting or in an out­patient surgery center.


Assisting with Surgical Procedures


Surgery performed in a medical office is restricted to the management of minor problems and injuries. The medical assistant is expected to assist with preparing the patient and setting up the sterile field. The following procedures must be used without exception when assisting with minor surgery. Individual facilities may have specific guidelines for some of these procedures; however, the theory behind sterile technique is universal, regardless of where you work.


Preparation of the Patient


Whether minor surgery is performed because of an unforeseen accident or is a planned, elective procedure, the patient needs both psychological and physical support. A patient facing a surgical procedure may be concerned about pain, disfigurement, and a possible diagnosis of cancer. An injured patient may feel anxious about medical bills or possible loss of employment. Because surgery is a frightening experience, the medical assistant must take the time, both preoperatively and at the time of surgery, to help the patient deal with fears and anxieties. The best way to help is to make sure that the patient understands the details of the procedure, that all questions are answered by the physician, and that the patient has the opportunity to talk about the procedure and voice any concerns.


Questions should be answered directly, but you should answer only the questions that are within your scope of knowledge and the policies of the office. If you cannot answer a question, assure the patient that you will relay it to the physician before the procedure and then be sure to do so. What may seem to be a minor or unimportant question to you may be a very frightening concern to the patient. The minor surgery room can be intimidating, so unless the patient is sedated, try to make conversation with him or her while you prepare for the physician’s arrival.


Preoperative preparation may include blood and urine tests, completion of a consent form, and gathering of the current history concerning any recent illnesses, medications, and allergies. Patient preparations before surgery may include a shave prep, cleansing enemas, food intake restrictions, special bathing, and administration of a sedative medication. On the day of surgery, the patient is instructed to empty the bladder and undress and gown as requested. The vital signs are recorded in preparation for the procedure.


Preoperative Instructions


When office surgery is planned, certain procedures are followed before the appointment. These include the following:



Informed Consent


The physician must have the patient’s written informed consent before beginning any surgical procedure. To sign an informed consent form permitting the physician to legally perform the surgery, the patient must understand what procedure will be performed, why it should be done, the potential risks and benefits of the surgery, alternative treatments (including no treatment), and the possible risks of any alternative treatment. This legal requirement is not met simply by having the patient sign an operative permit; a discussion must occur, during which the physician provides the patient or the patient’s legal representative with enough information to enable the person to decide whether to proceed with the proposed surgical treatment. After this discussion, the patient either consents to or refuses the surgery. The patient then signs or refuses to sign the consent form. If the patient signs with an X, the medical assistant should write “patient’s mark” beside the X and also should have a family member witness the signature. The discussion must be fully documented in the patient’s medical record. A copy of the signed form must also be included in the patient’s record. Treatment may not exceed the scope of the consent form.


The patient must not be under the influence of any sedative medication at the time he or she signs the consent form. This condition must never be violated.



57-4


Critical Thinking Application


Melissa is preparing a patient for a breast biopsy. The consent form has been signed and is in the chart. While Melissa is chatting with the patient during completion of the final setup for the procedure, it becomes clear the patient thinks she is having a “cyst removed” from her breast. What action should Melissa take, if any? What is the significance of what the patient said in this situation?

Stay updated, free articles. Join our Telegram channel

Apr 6, 2017 | Posted by in MEDICAL ASSISSTANT | Comments Off on Surgical Asepsis and Assisting with Surgical Procedures

Full access? Get Clinical Tree

Get Clinical Tree app for offline access