9: Circulation Procedures

Section Nine Circulation Procedures





PROCEDURE 48 Positioning the Hypotensive Patient



Maureen T. Quigley, MS, ARNP


Also known as modified Trendelenburg position.








PROCEDURE 49 Doppler Ultrasound for Assessment of Blood Pressure and Peripheral Pulses



Maureen T. Quigley, MS, ARNP




CONTRAINDICATIONS AND CAUTIONS




1. Use of an ultrasonic transmission gel recommended by the vendor assists in optimal sound transmission and protects the crystals, which are found in the probe. The crystals transmit and receive ultrasonic waves. In emergency situations, any surgical jelly or lubricant may be substituted for a conductive gel but ECG paste or cream should never be used as it may damage the probe (Parks Medical Electronics, 2005).


2. The probe should be checked regularly for damage to the electrode and integrity of the crystals.


3. Improper probe placement may lead to erroneous interpretations. Care should be taken to verify that the signal is coming from the intended vessel and not from a collateral vessel. This can be determined by assessing the quality of the sound, as described in the Complications section of this procedure.


4. Excess pressure on the probe may compress the artery and abolish the signal.


5. Verify sensitivity when signals are absent from a position where they would normally be expected. Sensitivity may be verified by checking one’s own pulses with the Doppler device.


6. The presence of a signal does not always indicate that circulation and perfusion are adequate to maintain viable tissue, just as absence of a signal does not always indicate that there is no blood flow through the vessel.


7. Tissue penetration varies, depending on which Doppler probe is used. A high-frequency (8 to 10 Hz) probe is usually used on the surface vessel sites; this probe is typically long and narrow. A low-frequency probe is usually used for deeper tissue sites, such as fetal heart tones; this probe is typically short and wide.


8. Falsely elevated pressures may be observed in patients with diabetes, obesity, or calcified vessels (Gorgas, 2004).




PROCEDURAL STEPS



Blood Pressure Measurement




1. Place the blood pressure cuff on the upper arm, the thigh, or the ankle and apply a transmission gel to the skin over the brachial-popliteal artery or the posterior-tibial artery. Be sure the cuff is high enough on the limb that the Doppler probe can access the area with the strongest pulse.


2. Turn on the Doppler instrument and turn down the volume. Insert the stethoscope earpieces, if applicable.


3. Adjust the volume control as necessary.


4. Identify the brachial-popliteal pulse or the posterior-tibial pulse with the Doppler instrument.


5. Position the probe over the artery and tilt it so that it is at a 45-degree angle along the length of the vessel to optimize frequency shifts and signal amplitude (Figure 49-2).


6. Inflate the blood pressure cuff until the arterial sounds are no longer audible.


7. Deflate the cuff slowly, listening for the first sound, which reflects the systolic pressure. Diastolic pressure is recorded at the point at which there is a decrease in arterial wall motion (Pickering, 2002).


8. Clean the gel from the patient’s skin with a wet towel or tissue.


9. Clean the face of the Doppler probe with a soft tissue. Do not use alcohol or other organic solvents to clean the probe. If the gel has dried on the probe, clean the probe with warm (not hot) running tap water but do not immerse the probe. The probe may be gas (ethylene oxide) sterilized at the lowest temperature consistent with good sterilization not to exceed 60° C [140° F]). A sterile water wash after sterilization is recommended (Parks Medical Electronics, 2005).








PROCEDURE 50 Measuring Postural Vital Signs



Margo E. Layman, MSN, RN, RNC, CN-A


Postural vital signs are also known as orthostatic vital signs and tilt test.




CONTRAINDICATIONS AND CAUTIONS




1. The value of orthostatic vital signs is disputed as there is no universal definition of how to perform them or what blood pressure and heart rate changes constitute “positive orthostatics.” Euvolemic patients may have orthostatic changes, whereas hypovolemic patients may not. Therefore, the vital signs should be interpreted in the context of the patient’s other signs and symptoms such as dizziness and visual dimming.


2. An assistant may be necessary because a patient with orthostatic hypotension may experience dizziness, lightheadedness, or syncope when moving from a lying to a standing position for postural vital sign measurement. Do not leave the patient alone during this procedure.


3. Orthostatic vital signs are contraindicated in patients with supine hypotension, shock, or a severe alteration in mental status, as well as in those who may have spinal, pelvic, or lower-extremity injuries (Gorgas, 2004).


4. Certain medications, such as sympatholytic drugs, diuretics, nitrates, narcotics, antihistamines, psychotropic agents, barbiturates, antihypertensives, and anticholergenics, can predispose a patient to orthostatic hypotension in the absence of hypovolemia. Studies have demonstrated a significant incidence of orthostatic hypotension, even in euvolemic patients (Irvin & White, 2004).


5. Paradoxical bradycardia may be observed in hypovolemic patients who have rapid and massive bleeding; this may be interpreted as orthostasis (Gorgas, 2004).


6. Prevent unreliable results by avoiding invasive or painful procedures during the measurement of postural vital signs.




PROCEDURAL STEPS




1. Measure the blood pressure and heart rate measurements after the patient has been in supine position for 2 to 3 minutes. Taking two sets of measurements and using the second set as baseline helps prevent false-positive results that are based on patient’s sympathetic response (Bradley & Davis, 2003).


2. Have the patient move from the supine to the sitting position (if three measurements are taken) or from supine to standing. If the patient is unable to stand for blood pressure measurement, try the high Fowler’s position, although the results may be less credible. A supine-to-standing measurement is more accurate than a supine-to-sitting measurement (Gorgas, 2004).


3. Question the patient about weakness, dizziness, or visual dimming associated with a change of position. Note any pallor or diaphoresis. These symptoms are as important as the measurement of vital signs. If the patient becomes extremely dizzy and needs to lie down or becomes syncopal, the measurement should be terminated.


4. Take the standing or sitting blood pressure (in the same arm as the initial readings) and the heart rate measurement within 1 minute. Support the patient’s forearm at heart level when taking the blood pressure to prevent an inaccurate measurement.


5. If an intermediate sitting measurement was taken, have the patient move into the standing position and repeat steps 3 and 4.


6. Return patient to supine or sitting position.


7. Note all measurements on the patient record, including the position in which they were taken (i.e., with the patient lying, sitting, or standing). Positive findings in adults are usually considered to be a heart rate increase of 30 beats/min, a decrease in systolic blood pressure of 20 mm Hg, a diastolic blood pressure decrease of 10 mm Hg, or symptoms of cerebral hypofusion, such as dizziness and syncope (Bradley & Davis, 2003). Other authors state that changes in blood pressure are too variable to be considered a reliable indicator of blood loss (Gorgas, 2004). The results of one study indicate that the two most valuable predictors for hypovolemia are a pulse increase of 30 beats/min or severe dizziness on standing (McGee, Abernethy, & Simel, 1999).







PROCEDURE 51 Pneumatic Antishock Garment



Reneé Semonin Holleran, RN, PhD, CEN, CCRN, CFRN, CTRN, FAEN


Pneumatic antishock garment is also known as PASG, military antishock garment (MAST), or shock pants.






PATIENT PREPARATION




1. If the mechanism of injury warrants immobilization, place the patient on a long spine board or a scoop stretcher (see Procedure 111). Position the garment on the board or the scoop stretcher before the patient. The patient may be log rolled onto the garment and the garment slid up under the patient (technique A), or, alternatively, the garment may be placed on the patient like trousers (technique B).


2. Remove all clothing and anything else below the waist to prevent compression injuries from objects such as belts when the pants are inflated.



PROCEDURAL STEPS



Inflation






Technique B


(This method is contraindicated in persons with suspected or confirmed spinal fractures.)







2. The abdominal compartment should be placed just below the rib cage so as not to reduce vital capacity and impair respiration.


3. Close the stopcock to the abdominal compartment and open the stopcocks to the leg compartments.


4. Attach the tubing from the foot pump to the three compartments. Some models have color-coded tubing for the different compartments.


5. Inflate the leg compartments with the pump until the pressure gauge (if present) indicates the appropriate amount of pressure. Some pants have a pop-off valve that limits the inflation pressure to 104 mm Hg. Crackling of the Velcro closures also indicates sufficient inflation. The goal is to achieve a systolic pressure of 100 mm Hg while using the lowest inflation pressure possible (less than 40 mm Hg). Patients in extremis may need rapid, simultaneous inflation of all three compartments to 100 mm Hg immediately on application of the garment. If the foot pump is not available, sufficient pressures may be obtained by manually inflating the compartments.


6. Recheck the patient’s vital signs. Inflation should be stopped when the systolic pressure reaches 100 mm Hg.


7. If the systolic pressure has not increased to more than 100 mm Hg, inflate the abdominal compartment. (Note: Pregnancy is a relative contraindication for inflation of the abdominal compartment.)


8. Assess vital signs and respiratory effort frequently.


9. Secure the patient to the long board or scoop stretcher as necessary.









PROCEDURE 52 Therapeutic Phlebotomy



Maureen T. Quigley, MS, ARNP


Therapeutic phlebotomy is also known as blood letting.






PROCEDURAL STEPS




1. Obtain an order including the diagnosis, most recent hemoglobin or hematocrit value, and the amount of blood to be removed.


2. Apply the tourniquet or blood pressure cuff and locate the most suitable antecubital vein. Remove the tourniquet or deflate the blood pressure cuff.


3. Cleanse the site with antiseptic solution.


4. Clamp the tubing with the hemostat at the patient’s end and insert the other end into the vacuum bottle. If using the Saf-T donor set, clamp the tubing to maintain the vacuum in the bottle and connect the 15-G stopper-piercing needle to the bottle.


5. Reapply the tourniquet or inflate the blood pressure cuff to a pressure between the patient’s systolic and diastolic readings.


6. Inject 1 to 2 ml of local anesthetic intradermally at the venipuncture site (optional).


7. Perform the venipuncture with the needle on the proximal end of the tubing.


8. Tape the needle in place and cover the site with an occlusive dressing.


9. Unclamp the hemostat and collect the desired amount of blood, usually 250 to 500 ml, in the collection bottle or bag.




10. Reclamp the tubing with the hemostat.


11. Release the tourniquet or deflate the cuff.


12. Withdraw the needle and apply pressure with a gauze dressing until the bleeding stops.


13. Assess and document the patient’s vital signs.







PROCEDURE 53 Pericardiocentesis



Andrew A. Galvin, APRN,BC, CEN


Pericardiocentesis is also known as pericardial tap.




CONTRAINDICATIONS AND CAUTIONS



Stay updated, free articles. Join our Telegram channel

Nov 8, 2016 | Posted by in NURSING | Comments Off on 9: Circulation Procedures

Full access? Get Clinical Tree

Get Clinical Tree app for offline access